The tracing is a clean normal sinus ECG. Cross a cable under Reversal, put noise on one electrode under Artifact, then try to remove it under Filters. Every lead an artifact reaches is flagged in orange on the paper.
Or test yourself: six tracings, each with one hidden fault to name.
Electrode to leads
RAI · II · aVR · aVL · aVF · V1–V6LAI · III · aVR · aVL · aVF · V1–V6LLII · III · aVR · aVL · aVF · V1–V6RLno lead (neutral)V1–V6its own lead only
The chest leads are measured against the Wilson central terminal, (RA+LA+LL)/3, so a faulty limb electrode reaches all six of them at one third strength.
Lead distribution
Readouts
Limb cable reversal
Chest cable swap
Limb electrode placement
Used for exercise and monitoring. The frontal axis moves rightward and I and aVL shrink; the chest leads barely change. Modelled here as an approximation: a 12° frontal rotation.
Where the noise enters
Mains hum with its harmonics. The notch rejects the fundamental and the harmonics together.
Breathing and slow movement. It is low frequency, so the high-pass removes it; a big swing pushes the trace off the paper.
Broadband noise, in µV RMS. A few tens of µV are tolerable on a 1 mV complex; above about 120 µV the P wave is buried.
Intermittent contact at one electrode: sudden steps, or the input hitting the rail. Each lead sees it with the sign and weight of that electrode in its own equation.
Presets
Raise it to remove baseline wander. Up to 0.5 Hz it is linear-phase, as on a diagnostic machine. Above 0.5 Hz it behaves like a monitoring filter and distorts the ST segment.
Removes mains interference at the frequency set under Artifact, and its harmonics.
Lower it to smooth muscle noise. Too low and the QRS loses height and its sharp detail.
Paper
A 1 mV calibration pulse sits at the left edge of every strip.
Signal source
Columns named I, II, III, aVR, aVL, aVF, V1 to V6, in mV or µV. I, II and V1 to V6 are needed; the other four are derived.
Vector loops
The single moving dipole behind the synthetic tracing, in the frontal and horizontal planes. Every lead is one projection of it.
Find the fault
Six tracings, one hidden fault each
The simulator sets a fault and hides the controls. Read the paper and name it. The orange flags and the explanation appear only after you answer.
Is it artifact? A bedside sequence
Is the rate physiologically possible?Above 300 a minute: artifact.
Is the pulse oximeter or arterial trace normal?Normal and regular: artifact.
Do real P, QRS and T march through the disturbance?Yes: artifact, with the true rhythm underneath.
Is there a fixed deflection just after every QRS?Yes: a mechanical (arterial pulse-tapping) artifact.
Which leads carry it?All leads: equipment or environment. One limb lead spared: the limb electrode it does not use. One chest lead: that chest electrode.
Never treat the ECG before you have checked the patient.
About this simulator
Education and training only. The tracing is synthetic. Do not use it for diagnosis or any patient-care decision.
What is exact. Lead formation (Einthoven and Goldberger), cable reversals, and the way a fault at one electrode is distributed to the twelve leads through the Wilson central terminal are computed from the lead equations. Every artifact is added at an electrode, never painted onto a lead.
What is simplified. The heartbeat is one cardiac dipole projected through the Dower transform, so it is a tidy normal sinus beat. The right-leg electrode is treated as a passive neutral at near left-leg potential. Mason-Likar placement is an approximation. Contact artifacts are shown after the filters, since no filter recovers a lost contact.
Mondal S, Raja DP, Muslim NA, Prabhu MA. Electrocardiographic artifacts in clinical practice: a logical approach to recognition and prevention. World J Cardiol 2026;18(3):116299.
Dower GE, Machado HB, Osborne JA. On deriving the electrocardiogram from vectorcardiographic leads. Clin Cardiol 1980;3:87-95.